![]()
CHAPTER 221. X-RAYS IN THE HEALING ARTS
GENERAL PROVISIONS Sec.
221.1. Purpose and scope.
221.2. Definitions.
221.3. [Reserved].
ADMINISTRATIVE CONTROLS
221.11. Registrant responsibilities.
221.12. Records, maintenance and associated information.
221.13. Information to be submitted by persons requesting approval to conduct healing arts screening.
221.14. [Reserved].
221.15. Use of X-rays in research on humans.
DIAGNOSTIC INSTALLATIONS GENERAL REQUIREMENTS
221.21. Diagnostic equipment requirements.
221.22. Battery charge indicator.
221.23. Leakage radiation from the diagnostic source assembly.
221.24. Radiation from components other than the diagnostic source assembly.
221.25. Beam quality.
221.26. Multiple tubes.
221.27. Mechanical support of tube head.
221.28. Technique indicators.
221.29. Kilovoltage (kV) accuracy.
221.30. Exposure reproducibility.
221.31. [Reserved].
221.31a. Locks.
221.32. [Reserved].
221.32a. Radiographic beam limitation.
221.33. [Reserved].
221.33a. Radiation from capacitor energy storage equipment in standby status.
221.34. [Reserved].
221.34a. Radiation exposure control.
221.35. [Reserved].
221.35a. Fluoroscopic X-ray systems.
221.36. [Reserved].
221.36a. Limitation of useful beam of fluoroscopic equipment.
221.37. [Reserved].
221.37a. Activation of fluoroscopic tube.
221.38. [Reserved].
221.38a. Entrance exposure rate.
221.39. [Reserved].
221.39a. Barrier transmitted radiation rate limits.
221.40. [Reserved].
221.40a. Indication of tube voltage and current.
221.41. [Reserved].
221.41a. Fluoroscopic timer.
221.42. [Reserved].
221.42a. Control of scattered radiation.
221.43. [Reserved].
221.43a. Mobile fluoroscopes.
221.44221.49 [Reserved].
INTRAORAL DENTAL RADIOGRAPHIC SYSTEMS
221.51221.55 [Reserved].
221.56. [Reserved].
OTHER SYSTEMS
221.61. Radiation therapy simulation systems.
221.62. [Reserved].
THERAPEUTIC X-RAY SYSTEMS WITH ENERGIES
LESS THAN 1 MEV
221.71. Equipment requirements.
221.72. Facility design requirements for systems capable of operating above 50 kVp.
221.73. Surveys.
221.74. Calibration.
221.75. Spot checks.
221.76. Operating procedures.
221.81221.102 [Reserved].
COMPUTED TOMOGRAPHY X-RAY SYSTEMS
221.201. Definitions.
221.202. Equipment requirements.
221.203. Facility design requirements.
221.204. Radiation measurements and performance evaluations.
221.205. Operating procedures.Authority The provisions of this Chapter 221 issued under section 301 of the The Atomic Energy Development and Radiation Control Act (73 P. S. § 1301) (Repealed), unless otherwise noted.
Source The provisions of this Chapter 221 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212, unless otherwise noted.
Cross References This chapter cited in 25 Pa. Code § 215.32 (relating to exemption qualifications); 25 Pa. Code § 225.104 (relating to X-ray detection systems for explosives, weapons and illegal items); 28 Pa. Code § 501.4 (relating to regulations); and 28 Pa. Code § 565.12 (relating to radiology service policy).
GENERAL
§ 221.1. Purpose and scope.
This chapter establishes requirements for the use of X-ray equipment by or under the supervision of a licensed practitioner of the healing arts. A registrant who uses X-rays in the healing arts shall comply with this chapter. This chapter is in addition to, and not in substitution for, other applicable provisions of this article.
Authority The provisions of this § 221.1 amended under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.1 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235. Immediately preceding text appears at serial pages (117772), (42615) and (4877) to (4880).
Cross References The provisions of this § 221.2 amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); and the Radon Certification Act (63 P. S. § § 20012014).
Source The provisions of this § 221.2 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823; amended May 16, 2008, effective May 17, 2008, 38 Pa.B. 2243. Immediately preceding text appears at serial pages (304479) to (304486).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems); and 25 Pa. Code § 221.201 (relating to definitions).
§ 221.3. [Reserved].
Source The provisions of this § 221.3 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved May 16, 2008, effective May 17, 2008, 38 Pa.B. 2243. Immediately preceding text appears at serial page (304486).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
ADMINISTRATIVE CONTROLS
§ 221.11. Registrant responsibilities.
(a) The registrant is responsible for directing the operation of X-ray systems under his administrative control and shall assure that the requirements of this article are met in the operation of the X-ray systems.
(b) An individual who operates an X-ray system shall be instructed adequately in the safe operating procedures and be competent in the safe use of the equipment. The instructions shall include items included in Appendix A (relating to determination of competence) and there shall be continuing education in radiation safety, biological effects of radiation, quality assurance and quality control.
(c) A chart, which specifies the techniques for examinations performed with the system, shall be provided in the vicinity of each diagnostic X-ray systems control panel. This chart shall include information pertinent to the particular examination, such as:
(1) The patients body part and anatomical size, or body part thickness, or age (for pediatrics), versus technique factors to be utilized.
(2) The type and size of the film or film-screen combination.
(3) The type of grid, if any.
(4) The type and location of placement of patient shielding-for example, gonad, and the like.
(5) For mammography, indication of kVp/target/filter combination.
(6) Source to image receptor distance to be used, except for dental intraoral radiography.
(d) Written safety procedures and rules shall be available at a facility including restrictions of the operating technique required for the safe operation of the particular X-ray system. The operator shall be able to demonstrate familiarity with the rules.
(e) Except for patients who cannot be moved out of the room, only the staff and ancillary personnel or other persons required for the medical procedure or training shall be in the room during the radiographic exposure. The following apply for individuals other than the patient being examined:
(1) Individuals shall be positioned so that no part of the body will be struck by the useful beam unless protected by at least 0.5 millimeter lead equivalent material. The lead equivalent of the material is to be determined at 60 kV.
(2) All persons required for the medical procedure shall be protected from the stray radiation by protective aprons or whole protective barriers of at least 0.25 millimeter lead equivalent or shall be so positioned that the persons are not in the direct line of the useful beam and the nearest portion of the body is at least 2 meters from both the tube head and the nearest edge of the image receptor.
(3) A patient who cannot be removed from the room shall be protected from the stray radiation by protective barriers of at least 0.25 millimeter lead equivalent material unless the shield would compromise the health of the individual or shall be so positioned that the patient is not in the direct line of the useful beam and the nearest portion of the body is at least 2 meters from both the tube head and the nearest edge of the image receptor.
(4) No individual, other than the patient being examined, may be in the useful beam, unless required to conduct the procedure.
(f) During diagnostic procedures in which the gonads are in the useful beam, gonad shielding of at least 0.5 millimeter lead equivalent shall be used for patients except for cases in which this would interfere with the diagnostic procedure.
(g) An individual may not be exposed to the useful beam except for healing arts purposes or under § 221.15 (relating to use of X-rays in research on humans). An exposure shall be authorized by a licensed practitioner of the healing arts. This provision specifically prohibits deliberate exposure for the following purposes:
(1) Exposure of an individual for training, demonstration or other nonhealing arts purposes.
(2) Exposure of an individual for the purpose of healing arts screening except as authorized by the Department. When requesting authorization, the registrant shall submit the information outlined in § 221.13 (relating to information to be submitted by persons requesting approval to conduct healing arts screening).
(h) If a patient or image receptor requires auxiliary support during a radiation exposure the following apply:
(1) Mechanical holding devices shall be used when the technique permits.
(2) The human holder shall be protected as required by subsection (e).
(3) An individual may not be used routinely to hold image receptors or patients.
(i) Procedures and auxiliary equipment designed to minimize patient and personnel exposure commensurate with the needed diagnostic information shall be utilized.
(j) The screen and film system used shall be spectrally compatible. Defective screens may not be used for diagnostic radiological imaging.
(k) With the exception of intraoral dental radiography, film may not be used without intensifying screens for routine diagnostic radiological imaging.
(l) The registrant shall have a quality assurance program. This quality assurance program shall be documented and be in accordance with guidelines established by the Department or by another appropriate organization recognized by the Department. At a minimum, the quality assurance program shall address repeat rate; image recording, processing and viewing; and maintenance and modifications to the quality assurance program. Records shall be maintained by the registrant for inspection by the Department for 3 years. The Departments guidelines and a list of recognized organizations will be maintained and made available on the Departments website and on request.
(m) Neither the X-ray tube housing nor the collimating device may be hand-held during the exposure.
Authority The provisions of this § 221.11 amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.11 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended November 16, 2001, effective November 17, 2001, 31 Pa.B. 6282; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceding text appears at serial pages (249282), (285667) to (285669).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems); and 25 Pa. Code § 221.42a (relating to control of scattered radiation).
§ 221.12. Records, maintenance and associated information.
The registrant shall maintain records of surveys, calibrations, maintenance and modifications performed on the X-ray systems including the names of persons who performed the services. The registrant shall keep these records for inspection by the Department for 5 years.
Authority The provisions of this § 221.12 amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.12 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial page (123681).
Cross References The provisions of this § 221.13 amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); amended under sections 301 and 302 of the Radiation Act (35 P. S. § § 7110.301 and 7110.302); section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); and the Radon Certification Act (63 P. S. § § 20012014).
Source The provisions of this § 221.13 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended November 16, 2001, effective November 17, 2001, 31 Pa.B. 6282; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823; amended May 16, 2008, effective May 17, 2008, 38 Pa.B. 2243. Immediately preceding text appears at serial pages (304489) to (304491).
Cross References This section cited in 25 Pa. Code § 221.11 (relating to registrant responsibilities); and 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.14. [Reserved].
Source The provisions of this § 221.15 issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.15 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3238. Immediately preceeding text appears at serial pages (249287) to (249288).
Cross References This section cited in 25 Pa. Code § 221.11 (relating to registrant responsibilities); and 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
DIAGNOSTIC INSTALLATIONS GENERAL REQUIREMENTS
§ 221.21. Diagnostic equipment requirements.
Diagnostic systems incorporating one or more certified components shall comply with 21 CFR 1020.301020.33.
Authority The provisions of this § 221.21 amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.21 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial page (123683).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.22. Battery charge indicator.
On battery-powered X-ray generators, the control panel shall have means to indicate visually whether the battery is adequately charged for proper operation.
Authority The provisions of this § 221.22 amended under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.22 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235. Immediately preceding text appears at serial page (4882).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.23. Leakage radiation from the diagnostic source assembly.
The leakage radiation from the diagnostic source assembly measured at a distance of 1 meter in any direction from the source may not exceed 100 milliroentgens (25.8 µC/kg) in 1 hour when the X-ray tube is operated at its leakage technique factors. Compliance shall be determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.
Authority The provisions of this § 221.23 amended under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.23 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235. Immediately preceding text appears at serial page (4882).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.24. Radiation from components other than the diagnostic source assembly.
The radiation emitted by a component other than the diagnostic source assembly may not exceed 2 milliroentgens (.516 µC/kg) in 1 hour at 5 centimeters from an accessible surface of the component when it is operated in an assembled X-ray system under conditions for which it was designed. Compliance shall be determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.
Authority The provisions of this § 221.24 amended under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.24 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235. Immediately preceding text appears at serial pages (4882) to (4883).
Cross References The provisions of this § 221.25 issued under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.25 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3238. Immediately preceeding text appears at serial pages (249290) and (285671).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.26. Multiple tubes.
If two or more radiographic tubes are controlled by one exposure switch, the tube or tubes which have been selected shall be clearly indicated prior to initiation of the exposure. This indication shall be both on the X-ray control panel and at or near the tube housing assembly which has been selected.
Authority The provisions of this § 221.26 issued under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.26 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235.
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.27. Mechanical support of tube head.
The tube housing assembly supports shall be adjusted so that the tube housing assembly will remain stable during an exposure unless tube housing movement is a designed function of the X-ray system.
Authority The provisions of this § 221.27 issued under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.27 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235.
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.28. Technique indicators.
(a) The technique factors for radiographic systems shall be indicated before exposure except for units utilizing automatic exposure controls, in which case the maximum mAs shall be indicated.
(b) The requirement of subsection (a) may be met by permanent markings on equipment having fixed technique factors. Indication of technique factors shall be visible from the operators position except in the case of spot films made by a fluoroscopist.
Authority The provisions of this § 221.28 issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.27 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial page (123686).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.29. Kilovoltage (kV) accuracy.
(a) For variable kV units, the kV output may not vary from the set-indicated value by more than 10% over the range of technique factors normally used. Discrepancies of more than 10% between set-indicated and measured kV values shall be investigated by a qualified expert or service agent and appropriate action taken.
(b) For fixed kV units, the kV output may not vary from the set-indicated value by more than 20% over the range of technique factors normally used. Discrepancies of more than 20% between set-indicated and measured kV values shall be investigated by a qualified expert or service agent and appropriate action taken.
Authority The provisions of this § 221.29 issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.29 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended November 16, 2001, effective November 17, 2001, 31 Pa.B. 6282; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceding text appears at serial pages (285672) to (285673).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.30. Exposure reproducibility for noncertified systems.
The coefficient of variation of exposure reproducibility may not exceed 0.10 when technique factors are held constant. This requirement shall be deemed to have been met when four exposures are made. This requirement applies when either manual techniques or automatic exposure control is used.
Authority The provisions of this § 221.30 issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); and the Radon Certification Act (63 P. S. § § 20012014).
Source The provisions of this § 221.30 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended May 16, 2008, effective May 17, 2008, 38 Pa.B. 2243. Immediately preceding text appears on serial page (304498).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.31. [Reserved].
Source The provisions of this § 221.31 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial pages (123686) to (123688).
§ 221.31a. Locks.
Position locking, holding and centering devices on X-ray systems shall function as intended.
Authority The provisions of this § 221.31a issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.31a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.32. [Reserved].
Source The provisions of this § 221.32a issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20) and the Radon Certification Act (63 P. S. § § 20012014).
Source The provisions of this § 221.32a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended November 16, 2001, effective November 17, 2001, 31 Pa.B. 6282; amended May 16, 2008, effective May 17, 2008, 38 Pa.B. 2243. Immediately preceding text appears at serial pages (304499) and (304500).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.33. [Reserved].
Source The provisions of this § 221.33 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial pages (123688) to (123689).
§ 221.33a. Radiation from capacitor energy storage equipment in standby status.
Radiation emitted from an X-ray tube when the exposure switch or timer in not activated may not exceed a rate of 2 milliroentgens (0.516 µC/kg) per hour at 5 centimeters from an accessible surface of a fully charged diagnostic source assembly, with the beam-limiting device fully open.
Authority The provisions of this § 221.33a issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.33a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended November 16, 2001, effective November 17, 2001, 31 Pa.B. 6282. Immediately preceding text appears at serial pages (249295) to (249296).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.34. [Reserved].
Source The provisions of this § 221.34a issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.34a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).
§ 221.35. [Reserved].
Source The provisions of this § 221.35 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial pages (123690) to (123691).
§ 221.35a. Fluoroscopic X-ray systems.
Fluoroscopic X-ray systems shall use an image intensifier and in addition to the requirements of § § 221.1221.34a, shall meet the requirements of § § 221.36a221.38a (relating to limitation of useful beam of fluoroscopic equipment; activation of fluoroscopic tube; and entrance exposure rate).
Authority The provisions of this § 221.35a issued under sections 301 and 302 of the Radiation Protection Act (31 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.35a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.
Cross References This section cited in 25 Pa. Code § 221.43a (relating to mobile fluoroscopes); and 25 Pa. Code § 221.61 (relating to radiation therapy simulation systems).
§ 221.36. [Reserved].
Source The provisions of this § 221.36a issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.36a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended November 16, 2001, effective November 17, 2001, 31 Pa.B. 6282; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceding text appears at serial pages (285678) to (285679).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems); and 25 Pa. Code § 221.43a (relating to mobile fluoroscopes).
§ 221.37. [Reserved].
Source The provisions of this § 221.37 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial pages (123691) to (123692).
§ 221.37a. Activation of fluoroscopic tube.
X-ray production in the fluoroscopic mode shall be controlled by a device which requires continuous pressure by the fluoroscopist for the entire time of the exposure (dead-man switch). When recording serial fluoroscopic images, the fluoroscopist shall be able to terminate X-ray exposures at any time, but means may be provided to permit completion of any single exposure of the series in process.
Authority The provisions of this § 221.37a issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.37a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems); 25 Pa. Code § 221.43a (relating to mobile fluoroscopes); and 25 Pa. Code § 221.61 (relating to radiation therapy simulation systems).
§ 221.38. [Reserved].
Source The provisions of this § 221.38 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial page (123692).
§ 221.38a. Entrance exposure rate.
(a) Fluoroscopic systems without high level control. The exposure rate may not exceed 10 roentgens (2.58 mC/kg) per minute except during recording of fluoroscopic images.
(b) Fluoroscopic systems with high level control.
(1) When the high level control is activated, the maximum exposure rate shall be 20 roentgens (5.16 mC/kg) per minute.
(2) When the high level control is not activated, the maximum exposure rate shall be 10 roentgens (2.58 mC/kg) per minute.
(3) Special means of activation of high level controls are required. The high level control shall only be operable when continuous manual activation is provided by the operator.
(4) There shall be an indication to the fluoroscopist that the high level control is being used.
(c) Frequency of output measurements. Output measurements to show compliance with this section shall be made at least annually and after maintenance that could affect the output of the machine.
(d) Compliance requirements. Compliance with subsections (a)(c) shall be determined as follows:
(1) If the source is below the table, the exposure rate shall be expressed for the center of the useful beam 1 centimeter above the tabletop or cradle with the image intensifier 30 centimeters above the tabletop or cradle.
(2) If the source is above the table, the exposure rate shall be measured at 30 centimeters above the tabletop with the end of the beam-limiting device or spacer positioned as closely as possible to the point of measurement.
(3) In a c-arm type of fluoroscope, the exposure rate shall be measured at 30 centimeters from the input surface of the fluoroscopic imaging assembly with the source at its closest possible position of operation.
(4) The tube potential and current shall be set to give the maximum exposure possible from the X-ray system. For systems with automatic exposure control, at least 3 millimeters of lead shall be placed between the measuring device and image receptor.
(5) The measurement shall be made at the center of the useful beam.
Authority The provisions of this § 221.38a issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.38a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823.
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems); and 25 Pa. Code § 221.43a (relating to mobile fluoroscopes).
§ 221.39. [Reserved].
Source The provisions of this § 221.39 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial pages (123692) to (123693).
§ 221.39a. Barrier transmitted radiation rate limits.
The protective barrier may not transmit more than 2 milliroentgens (.516 µmC/kg) per hour at 10 centimeters from an accessible surface of the fluoroscopic imaging assembly for each roentgen per minute of entrance exposure rate.
Authority The provisions of this § 221.39a issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.39a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.
Cross References This section cited in 25 Pa. Code § 221.43a (relating to mobile fluoroscopes).
§ 221.40. [Reserved].
Source The provisions of this § 221.40 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial page (123693).
§ 221.40a. Indication of tube voltage and current.
During fluoroscopy and cinefluorography, the voltage and the current shall be continuously indicated.
Authority The provisions of this § 221.40a issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.40a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.
Cross References This section cited in 25 Pa. Code § 221.43a (relating to mobile fluoroscopes); and 25 Pa. Code § 221.61 (relating to radiation therapy simulation systems).
§ 221.41. [Reserved].
Source The provisions of this § 221.41 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial page (123693).
§ 221.41a. Fluoroscopic timer.
A cumulative timing device activated by the fluoroscope switch shall be provided. It shall indicate the passage of a predetermined period of irradiation either by an audible signal or by temporary or permanent interruption of the irradiation when the increment of exposure time exceeds a predetermined limit not exceeding 5 minutes.
Authority The provisions of this § 221.41a issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.41a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems); 25 Pa. Code § 221.43a (relating to mobile fluoroscopes); and 25 Pa. Code § 221.61 (relating to radiation therapy simulation systems).
§ 221.42. [Reserved].
Source The provisions of this § 221.42 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial page (123694).
§ 221.42a. Control of scattered radiation.
(a) Fluoroscopic table designs when combined with normal operating procedures shall be of a type so no unprotected part of the staff or an ancillary individuals whole body is exposed to unattenuated scattered radiation which originates from under the table. The attenuation required may be not less than .25 millimeter lead equivalent.
(b) Equipment configuration when combined with normal operating procedures shall be of a type so that no portion of the staff or an ancillary individuals whole body, except the extremities, is exposed to the unattenuated scattered radiation emanating from above the table top unless one of the following criteria is met:
(1) The individual is at least 120 centimeters from the center of the useful beam.
(2) The radiation has passed through at least .25 millimeter of lead equivalent materialfor example, drapes, bucky-slot cover (film-tray cover panel), sliding or folding panel or self supporting curtainsin addition to lead equivalency provided by the protective apron referred to in § 221.11(e) (relating to registrant responsibilities).
Authority The provisions of this § 221.42a issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.42a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.
Cross References This section cited in 25 Pa. Code § 221.43a (relating to mobile fluoroscopes).
§ 221.43. [Reserved].
Source The provisions of this § 221.43 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial pages (123694) to (123695).
§ 221.43a. Mobile fluoroscopes.
In addition to the other requirements of § § 221.35a221.42a, mobile fluoroscopes shall provide image intensification.
Authority The provisions of this § 221.43a issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.43a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.
§ § 221.44221.49. [Reserved].
Source The provisions of these § § 221.44221.49 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial pages (123694) to (123698).
§ § 221.51221.55. [Reserved].
Source The provisions of these § § 221.51221.55 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial pages (123698) to (123701).
§ 221.56. [Reserved].
Source The provisions of this § 221.56 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved November 16, 2001, effective November 17, 2001, 31 Pa.B. 6282. Immediately preceding text appears at serial pages (249304) and (254511).
OTHER SYSTEMS
§ 221.61. Radiation therapy simulation systems.
(a) Fluoroscopic systems used solely for radiation therapy simulations shall comply with § § 221.35a, 221.37a, 221.40a and 221.41a. The requirements in § 221.41a (relating to fluoroscopic timer) may also be satisfied if a means is provided to indicate the cumulative time that an individual patient has been exposed to X-rays. In this case, procedures shall require that the timer be reset between examinations.
(b) CT units used solely for therapy simulations shall comply with § § 221.202(f)(1), (7) and (8) and 221.203 (relating to equipment requirements; and facility design requirements).
Authority The provisions of this § 221.61 issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.61 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended November 16, 2001, effective November 17, 2001, 31 Pa.B. 6282; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceding text appears at serial page (285683).
§ 221.62. [Reserved].
Source The provisions of this § 221.62 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial pages (123701) to (123702).
THERAPEUTIC X-RAY SYSTEMS WITH ENERGIES
LESS THAN 1 MEV
§ 221.71. Equipment requirements.
(a) When the tube is operated at its leakage technique factors, the leakage radiation may not exceed:
(1) One hundred milliroentgens (25.8µC/kg) per hour at 5 centimeters from the surface of the tube housing assembly for contact therapy systems.
(2) One roentgen (.258 mC/kg) per hour at 1 meter from the source for 0-150 kVp systems manufactured or installed prior to December 19, 1987.
(3) One hundred milliroentgens (25.8µC/kg) per hour at 1 meter from the source for 0-150 kVp systems manufactured on or after December 19, 1987.
(4) One roentgen (.258 mC/kg) per hour at 1 meter from the source for 151 to 500 kVp systems.
(5) One-tenth percent of the exposure rate of the useful beam 1 meter from the source for 501 to 999 kVp systems at 1 meter from the source.
(b) Fixed diaphragms or cones used for limiting the useful beam must provide at least the same protection as required by the tube housing assembly.
(c) Beam limiting devices may, for the portion of the useful beam blocked by these devices, transmit not more than 5% of the original X-ray beam intensity at the maximum voltage and maximum treatment filter. This requirement does not apply to auxiliary blocks or materials placed in the useful beam to shape the useful beam to the individual patient.
(d) The filter system shall be designed so that:
(1) Filters cannot be accidentally displaced from the useful beam at any possible tube orientation.
(2) The radiation at 5 centimeters from the filter insertion slot opening does not exceed 30 roentgens (7.74 mC/kg) per hour under operating conditions.
(3) A filter is marked as to its material of construction and its thickness. For wedge filters, the wedge factor and wedge angle shall appear on the wedge or wedge tray.
(4) On equipment purchased after January 1, 1971, a filter indication system shall be used on therapy machines using changeable filters. The system must indicate from the control panel the presence or absence of a filter and be designed to permit easy recognition of an added filter in place.
(5) An X-ray system equipped with a beryllium or other low-filtration window shall be clearly labeled as such upon the tube housing assembly and at the control panel.
(e) The tube housing assembly shall be immobilized during stationary treatments.
(f) The tube housing assembly shall be so marked that it is possible to determine the location of the focal spot to within 5 millimeters, and the marking shall be readily accessible for use during calibration procedures.
(g) Contact therapy tube housing assemblies shall have a removable shield of at least .5 millimeter lead equivalency at 100 kVp that can be positioned over the entire useful beam exit port during periods when the beam is not in use.
(h) Systems of greater than 150 kVp manufactured after December 19, 1987, must have a beam monitor system which meets the following requirements:
(1) Not allow irradiation until a preselected value of exposure has been made at the treatment control panel.
(2) Independently terminate irradiation when the preselected exposure has been reached.
(3) Be designed so that, in the event of a system malfunction or electrical power failure or other interruption, the dose administered to a patient prior to the interruption can be accurately determined.
(4) Have a control panel display which maintains the reading until intentionally reset to zero.
(5) Have a control panel display which does not have scale multiplying factors and utilizes a design so that increasing dose is displayed by increasing numbers.
(i) The following apply to timers on the equipment:
(1) A timer shall be provided which has a display at the control panel. The timer must be graduated in minutes and fractions of minutes. The timer must have a preset time selector and an elapsed time indicator.
(2) The timer must be a cumulative timer which activates with the radiation and retains its reading after irradiation is interrupted or terminated. After irradiation is terminated and before irradiation can be reinitiated, it shall be necessary to reset the timer to zero.
(3) The timer must terminate irradiation when a preselected time has elapsed if a dose monitoring system present has not previously terminated irradiation.
(4) The timer must permit accurate presetting and determination of exposure time as short as 1 second.
(5) The timer may not permit an exposure if set at zero.
(6) The timer may not activate until the shutter is opened when patient irradiation is controlled by a shutter mechanism.
(j) The control panel, in addition to the displays required in this section, must have:
(1) An indication of power status.
(2) An indication of X-ray production.
(3) The means of indicating X-ray tube current and voltage.
(4) The means of terminating an exposure.
(k) When a control panel may energize more than one X-ray tube, the following requirements shall be met:
(1) It must be possible to activate only one X-ray tube at one time.
(2) There must be an indication at the control panel identifying which X-ray tube is energized.
(3) There must be an indication at the tube housing assembly when that tube is energized.
(l) There must be a means of determining the SSD to within 5 millimeters.
(m) Unless it is possible to bring the X-ray output to the prescribed exposure parameters within 5 seconds, the entire useful beam shall be automatically attenuated by a shutter having a lead equivalency not less than that of the tube housing assembly.
(1) After the unit is at operating parameters, the shutter shall be controlled electrically by the operator from the control panel.
(2) An indication of shutter position must appear at the control panel.
Authority The provisions of this § 221.71 issued under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); and the Radon Certification Act (63 P. S. § § 20012014).
Source The provisions of this § 221.71 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended May 16, 2008, effective May 17, 2008, 38 Pa.B. 2243. Immediately preceding text appears at serial pages (304510) to (304512).
§ 221.72. Facility design requirements for systems capable of operating above 50 kVp.
(a) Provision shall be made to permit continuous observation of and communication with the patient during irradiation.
(b) Windows, mirror systems or closed-circuit television viewing screens used for observing the patient shall be so located that the operator can maintain direct surveillance over both the control panel and the patient.
(c) Treatment rooms which contain an X-ray system capable of operating above 150 kVp shall meet the following additional requirements:
(1) Necessary shielding, except for a beam interceptor, shall be provided by fixed barriers.
(2) The control panel shall be outside the treatment room or in a shielded booth.
(3) Doors of the treatment room shall be electrically interlocked to the control panel so that X-ray production cannot occur unless entrance doors are closed.
(4) Interlocks shall be provided so that, when a door of the treatment room is opened, either the machine will shut off automatically or the radiation level within the room will be reduced to an average of not more than two milliroentgens (.52 µC/kg) per hour and a maximum of ten milliroentgens (2.58 µC/kg) per hour at a distance of 1 meter in any direction from the target; or interlocks shall energize a conspicuous visible or audible alarm signal so that the individual entering and the operator are made aware of the entry. After a shut-off or reduction in output, it shall be possible to restore the machine to full operation only from the control panel.
(5) Treatment room entrances shall be provided with warning lights, which will indicate when the useful beam is on, in a readily observable position near the outside of access doors.
Authority The provisions of this § 221.72 issued under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.72 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235.
§ 221.73. Surveys.
(a) A facility shall have a survey made by, or under the direction of, a qualified expert. The survey shall also be done after a change in the facility or equipment which might cause a change in radiation levels.
(b) The qualified expert or radiological physicist shall report the survey results in writing to the individual in charge of the facility and a copy of the report shall be maintained by the registrant for inspection by the Department. The facility shall be operated in compliance with limitations indicated by the survey.
Authority The provisions of this § 221.73 issued under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.73 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceeding text appears at serial page (249309).
§ 221.74. Calibration.
(a) The calibration of an X-ray system shall be performed at intervals not to exceed 1 year and after a change of replacement of components which could cause a change in the radiation output.
(b) The calibration of the radiation output of the X-ray system shall be performed by or under the direction of a qualified expert for radiation therapy calibration who is physically present at the facility during the calibration.
(c) The calibration of the radiation output of an X-ray system shall be performed with a calibrated instrument. The calibration of the instrument shall be traceable to a National standard. The instrument shall have been calibrated within the preceding 2 years.
(d) Calibrations made under this section shall be made so that the dose at a reference point in soft tissue may be calculated as accurately as possible but with an uncertainty of no greater than 5%.
(e) The calibration of the X-ray system shall include, but is not limited to, the following determinations:
(1) The exposure rates for each combination of field size, technique factors, filter and treatment distance used.
(2) The degree of congruence between the radiation field and the field indicated by the localizing device if a device is present.
(3) An evaluation of the uniformity of the largest radiation field used.
(f) Records of calibration performed under this section shall be maintained by the registrant for at least 5 years after completion of the calibration.
(g) A copy of the most recent X-ray system calibration shall be available at the control panel.
Authority The provisions of this § 221.74 issued under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.74 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceeding text appears at serial pages (249309) to (249310).
Cross References The provisions of this § 221.75 issued under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.75 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceeding text appears at serial pages (249310) to (249311).
§ 221.76. Operating procedures.
(a) Therapeutic X-ray systems shall be secured to prevent unauthorized use whenever the system is unattended.
(b) If a patient is held in position for radiation therapy, mechanical supporting or restraining devices shall be used.
(c) The tube housing assembly may not be held by an individual during exposures.
(d) No individual other than the patient may be in the treatment room during irradiation unless protected by a shielded booth.
(e) Interlocks, on-off beam control mechanisms and safety and warning devices shall be checked and appropriately serviced at least once in a calendar year.
Authority The provisions of this § 221.76 issued under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.76 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235.
§ § 221.81221.102. [Reserved].
Source The provisions of these § § 221.81221.102 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; reserved October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894. Immediately preceding text appears at serial pages (123709) to (123726).
COMPUTED TOMOGRAPHY X-RAY SYSTEMS
§ 221.201. Definitions.
In addition to the definitions in § § 215.2 and 221.2 (relating to definitions), the following words and terms when used in this section and § § 221.202221.205, have the following meanings, unless the context clearly indicates otherwise:
CSContrast scaleThe change in the linear attenuation coefficient per CT number relative to water; that is:CS = (Ux-Uw)/((CT)x-(CT)w)
Where:
Ux = Linear attenuation coefficient of the material of interest
Uw = Linear attenuation coefficient of water
(CT)x = CT number of the material of interest
(CT)w = CT number of water
CT numberThe number used to represent the X-ray attenuation associated with each elemental area of the CT image.
CTComputed tomographyThe production of a tomogram by the acquisition and computer processing of X-ray transmission data.
CTDIComputed tomography dose indexThe integral of the dose profile along a line perpendicular to the tomographic plane divided by the product of the nominal tomographic section thickness and the number of tomograms produced in a single scan.
CT conditions of operationThe selectable parameters governing the operation of a CT X-ray system including, but not limited to, nominal tomographic section thickness, filtration and the technique factors as defined in this chapter.
DetectorA device that provides a signal or other indication suitable for measuring one or more quantities of incident radiation.
Elemental areaThe smallest area within a tomogram for which the X-ray attenuation properties of a body are depicted.
GantryThe tube housing assemblies, beam-limiting devices, detectors, transformers, if applicable, and the supporting structures and frames which hold these components.
LuxA unit illumination equivalent to 1 lumen per square centimeter or .0929 foot-candles.
MSADMultiple scan average doseThe calculated average dose to the tissue within each slice in a series utilizing an ion chamber. The MSAD is calculated using the following equation:MSAD = (F x K x L x E)/(T x N)
Where
F = Factor to convert exposure in air to absorbed dose in lucite in RADS/mR
K = Calibration factor to account for the ion chambers response and volume.
L = Effective length of ion chamber in millimeters (mm)
E = Exposure reading in milliroentgen (mR)
T = Nominal slice thickness in millimeters (mm) and
N = Number of slices per scan
Multiple tomogram systemA computed tomography X-ray system which obtains X-ray transmission data simultaneously during a single scan to produce more than one tomogram.
NoiseThe standard deviation of the fluctuations in the CT number expressed as a percentage of the attenuation coefficient of water. Its estimate (Sn) is calculated using the following expression:Sn = 100 x CS x S/Uw
Where:
CS = Contrast scale
Uw = Linear attenuation coefficient of water.
S = Estimated standard deviation of the CT number of picture elements in a specified area of the CT image.
Nominal tomographic section thicknessThe full-width at half-maximum of the sensitivity profile taken at the center of the cross-sectional volume over which X-ray transmission data are collected.
Performance phantomA phantom which has a capability of providing an indication of contrast scale, noise, nominal tomographic section thickness, the resolution capability of the CT system for low and high contrast objects, and measuring the mean CT number for water or other reference materials.
Picture elementSee elemental area.
PixelSee elemental area.
Reference planeA plane which is at a known fixed distancewhich could be zeroto the tomographic plane and parallel to it.
ScanThe complete process of collecting X-ray transmission data for the production of a tomogram. Data may be collected simultaneously during a single scan for the production of one or more tomograms.
Scan incrementThe amount of relative displacement of the patient with respect to the CT X-ray system between successive scans measured along the direction of the displacement.
Scan sequenceA preselected set of two or more scans performed consecutively under preselected CT conditions of operation.
Scan timeThe period of time between the beginning and end of X-ray transmission data accumulation for a single scan.
Sensitivity profileThe relative response of the CT X-ray system as a function of position along a line perpendicular to the tomographic plane.
Single tomogram systemA CT X-ray system which obtains X-ray transmission data during a scan to produce a single tomogram.
Technique factorsThe conditions of operation, specified as follows:(i) For CT equipment designed for pulsed operations, peak tube potential, scan time in seconds, X-ray pulse width in seconds and the number of X-ray pulses per second or per mAs.
(ii) For CT equipment not designed for pulsed operation, peak tube potential, and either tube current and scan time in seconds or the product of tube current and exposure time in mAs.
TomogramThe depiction of the X-ray attenuation properties of a section through a body.
Tomographic planeThe geometric plane which is identified as corresponding to the output tomogram.
Tomographic sectionThe volume of an object whose X-ray attenuation properties are imaged in a tomogram.
Authority The provisions of this § 221.201 issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); and the Radon Certification Act (63 P. S. § § 20012014).
Source The provisions of this § 221.201 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended May 16, 2008, effective May 17, 2008, 38 Pa.B. 2243. Immediately preceding text appears at serial pages (304515) to (304518).
Cross References The provisions of this § 221.202 issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.202 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended November 16, 2001, effective November 17, 2001, 31 Pa.B. 6282; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceding text appears at serial pages (285686) to (285688).
Cross References This section cited in 25 Pa. Code § 221.61 (relating to radiation therapy simulation systems); 25 Pa. Code § 221.201 (relating to definitions); and 25 Pa. Code § 221.204 (relating to radiation measurements and performance evaluations).
§ 221.203. Facility design requirements.
(a) Oral communication. Provision shall be made for oral communication between the patient and the operator at the control panel.
(b) Viewing systems.
(1) A means shall be provided to permit continuous observation of the patient during irradiation and shall be located so that the operator can observe the patient from the control panel.
(2) If the primary viewing system is by electronic means, an alternate viewing system, which may be electronic, shall be available for use in the event of failure of the primary viewing system.
Authority The provisions of this § 221.203 issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.203 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.
Cross References The provisions of this § 221.204 issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.204 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceeding text appears at serial pages (285688) and (249317) to (249318).
Cross References The provisions of this § 221.205 issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.205 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceeding text appears at serial pages (249318) to (249319).
Cross References This section cited in 25 Pa. Code § 221.201 (relating to definitions).
APPENDIX A
DETERMINATION OF COMPETENCE
The registrant shall ensure that individuals who operate diagnostic X-ray equipment have received training on the subjects listed in this appendix. The individual shall be trained and competent in the general operation of the X-ray equipment, and in the following subject areas, as applicable to the procedures performed and the specific equipment utilized:
(1) Basic properties of radiation.
(2) Units of measurement.
(3) Sources of radiation exposure.
(4) Methods of radiation protection.
(5) Biological effects of radiation exposure.
(6) X-ray equipment.
(7) Image recording and processing.
(8) Patient exposure and positioning.
(9) Procedures.
(10) Quality assurance.
(11) Regulations.
Authority The provisions of this Appendix A issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this Appendix A adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceeding text appears at serial page (249320).
Cross References This appendix cited in 25 Pa. Code § 221.11 (relating to registrant responsibilities).
No part of the information on this site may be reproduced for profit or sold for profit.
This material has been drawn directly from the official Pennsylvania Code full text database. Due to the limitations of HTML or differences in display capabilities of different browsers, this version may differ slightly from the official printed version.