Pennsylvania Code & Bulletin
COMMONWEALTH OF PENNSYLVANIA

• No statutes or acts will be found at this website.

The Pennsylvania Code website reflects the Pennsylvania Code changes effective through 54 Pa.B. 488 (January 27, 2024).

Pennsylvania Code



Subchapter D. ELECTRONIC DEVICES (NONRADAR) WHICH
MEASURE ELAPSED TIME BETWEEN TWO SENSORS


Sec.


105.51.    [Reserved].
105.52.    [Reserved].
105.52a.    Application of subchapter.
105.53.    Operation of maintenance and calibration stations.
105.54.    Calibration forms.
105.55.    Testing forms.
105.56.    Manner of calibration and testing.

Notes of Decisions

   Since the former PennDot regulation, 67 Pa. Code §  337.25, classifying ESP Machine (utilizes two sensors taped across street, connected to processing and display unit, to measure vehicle speed) as an electrical or mechanical device, rather than electronic device was an interpretive rather than legislative regulation, the Court was not bound thereby, the court concluded that the ESP Machine was an electronic device which could not be used by local police. Commonwealth v. DePasquale, 501 A.2d 626 (Pa. 1985).

§ 105.51. [Reserved].


Source

   The provisions of this §  105.51 adopted June 23, 1978, effective June 24, 1978, 8 Pa.B. 1692; amended December 15, 1978, effective December 16, 1978, 8 Pa.B. 3578; amended May 18, 1979, effective May 19, 1979, 9 Pa.B. 1632; republished and compiled August 1, 1980, effective August 2, 1980, 10 Pa.B. 3199; amended November 12, 1982, effective November 13, 1982, 12 Pa.B. 3946; reserved March 30, 1984, effective March 31, 1984, 14 Pa.B. 1089. Immediately preceding text appears at serial page (83723).

§ 105.52. [Reserved].


Source

   The provisions of this §  105.52 adopted June 23, 1978, effective June 24, 1978, 8 Pa.B. 1692; amended December 15, 1978, effective December 16, 1978, 8 Pa.B. 3578; amended September 28, 1979, effective September 29, 1979, 9 Pa.B. 3324; amended January 11, 1980, effective January 12, 1980, 10 Pa.B. 131; amended May 10, 1980, effective May 17, 1980, 10 Pa.B. 1980; amended May 30, 1980, effective May 31, 1980, 10 Pa.B. 2162; republished and compiled August 1, 1980, effective August 2, 1980, 10 Pa.B. 3199; amended October 17, 1980, effective October 18, 1980, 10 Pa.B. 4101; amended January 15, 1982, effective January 16, 1982, 12 Pa.B. 299; amended May 28, 1982, effective May 29, 1982, 12 Pa.B. 1698; amended November 12, 1982, effective November 13, 1982, 12 Pa.B. 3946; amended November 19, 1982, effective November 20, 1982, 12 Pa.B. 3983; amended September 2, 1983, effective September 3, 1983, 13 Pa.B. 2705; amended December 30, 1983, effective December 31, 1983, 13 Pa.B. 4021; reserved March 30, 1984, effective March 31, 1984, 14 Pa.B. 1089. Immediately preceding text appears at serial pages (83723) to (83724).

§ 105.52a. Application of subchapter.

 This subchapter governs the calibrating and testing of electronic devices—nonradar—which measure elapsed time between two sensors, for use by a police officer, under 75 Pa.C.S. §  3368(c)(3) (relating to speed timing devices).

Source

   The provisions of this §  105.52a adopted August 22, 1986, effective August 23, 1986, 16 Pa.B. 3119; readopted December 4, 1987, effective December 5, 1987, 17 Pa.B. 5051; amended February 17, 1989, effective February 18, 1989, 19 Pa.B. 640. Immediately preceding text appears at serial page (122758).

Cross References

   This section cited in 67 Pa. Code §  105a.4 (relating to system and speed timing device use and certification).

§ 105.53. Operation of maintenance and calibration stations.

 (a)  Maintenance and calibration. Maintenance and calibration of electronic devices governed by this subchapter shall be provided, as necessary, upon request by local police.

 (b)  Hours. Stations for the maintenance and calibration of electronic devices shall be operated during a regular work week, Monday through Friday during the hours of 8 a.m. to 5 p.m.

 (c)  Personnel. A station for the maintenance and calibration of electronic devices shall employ on a full-time basis:

   (1)  An electrical engineer or metrologist.

   (2)  An associate engineer, associate metrologist or technician.

 (d)  Presumption of accuracy. Either the electrical engineer, associate engineer, metrologist, associate metrologist or technician shall be present at the time the test is conducted. The certification that the electronic speed-timing device has been tested and found to be accurate shall create a presumption that the requirements of this subsection have been fulfilled.

 (e)  Damage to or loss of device. The station for the maintenance and calibration of electronic devices is responsible for damages or losses of electronic devices while they are being serviced or calibrated.

 (f)  Malfunctioning devices. A malfunctioning electronic device shall be returned to the manufacturer or to a service center approved by the manufacturer and tagged with the date tested, the serial number of the device and a notation of the general condition of the device.

 (g)  Location. Electronic testing and repair shall take place at an approved electronic testing station, except that, upon approval of the Department, an electronic testing station may operate mobile test units of a type approved by the Department.

Source

   The provisions of this §  105.53 adopted June 23, 1978, effective June 24, 1978, 8 Pa.B. 1692; republished and compiled August 1, 1980, effective August 2, 1980, 10 Pa.B. 3199; amended August 22, 1986, effective August 23, 1986, 16 Pa.B. 3119; readopted December 4, 1987, effective December 5, 1987, 17 Pa.B. 5051; amended February 17, 1989, effective February 18, 1989, 19 Pa.B. 640. Immediately preceding text appears at serial page (122758).

Cross References

   This section cited in 67 Pa. Code §  105a.4 (relating to system and speed timing device use and certification).

§ 105.54. Calibration forms.

 Certificates of accuracy, provided by the Department, shall be filed for electronic devices calibrated and tested as follows:

   (1)  Original and duplicate shall accompany electronic device.

   (2)  One duplicate copy shall be retained by the maintenance and calibration station for at least 2 years.

   (3)  A duplicate copy, executed and signed in the same manner as the original, has the same force and effect as the original.

Source

   The provisions of this §  105.54 adopted June 23, 1978, effective June 24, 1978, 8 Pa.B. 1692; republished and compiled August 1, 1980, effective August 2, 1980, 10 Pa.B. 3199; amended August 22, 1986, effective August 23, 1986, 16 Pa.B. 3119; readopted December 4, 1987, effective December 5, 1987, 17 Pa.B. 5051; amended February 17, 1989, effective February 18, 1989, 19 Pa.B. 640. Immediately preceding text appears at serial pages (122758) to (122759).

Cross References

   This section cited in 67 Pa. Code §  105a.4 (relating to system and speed timing device use and certification).

§ 105.55. Testing forms.

 Technical data forms, provided by the Department, shall be filed for electronic devices tested as follows:

   (1)  One original shall accompany the electronic device.

   (2)  One duplicate copy shall be retained by the maintenance and calibration station for at least 2 years.

   (3)  A duplicate copy, executed and signed in the same manner as the original, has the same force and effect as the original.

Source

   The provisions of this §  105.55 adopted June 23, 1978, effective June 24, 1978, 8 Pa.B. 1692; republished and compiled August 1, 1980, effective August 2, 1980, 10 Pa.B. 3199; amended August 22, 1986, effective August 23, 1986, 16 Pa.B. 3119; readopted December 4, 1987, effective December 5, 1987, 17 Pa.B. 5051; amended February 17, 1989, effective February 18, 1989, 19 Pa.B. 640. Immediately preceding text appears at serial page (122759).

Cross References

   This section cited in 67 Pa. Code §  105a.4 (relating to system and speed timing device use and certification).

§ 105.56. Manner of calibration and testing.

 (a)  Electrical Speed-Timing System—Manufactured by Richard Hageman, 98 South Penn Dixie Road, Nazareth, Pennsylvania. Calibration requirements are as follows:

   (1)  Required equipment. The following equipment or equivalent substitute equipment is required for calibration:

     (i)    USM 207 Universal Counter.

     (ii)    STALO E1010V.

     (iii)   Miscellaneous test leads.

   (2)  Calibration procedure. (See Appendix A, Figure 3 for equipment set up.) The calibration procedure shall be as follows:

     (i)   Set counter (USM 207) to Time (B-C) Mode.

     (ii)   Throw start switch on USM 207 and Electrical Speed-Timing System will start at same time.

     (iii)   Switch start override on.

     (iv)   Throw stop switch, stopping USM 207 and Electrical Speed-Timing System at same time.

     (v)   Read and record elapsed time of both devices.

     (vi)   Miles per hour (mph) is calculated as follows:

       (A)   Feet per second = measured distance—in feet—divided by time elapsed—in seconds.

       (B)   mph = .6818 x feet per second.

     (vii)   If the readout on the face of the Electrical Speed-Timing System differs by more than 1 mph, the unit shall be returned to the manufacturer and tagged with the date tested, the serial number of the device and a notation of the general condition of the device.

 (b)  Model TK 100, Excessive Speed Preventor (ESP)—Manufactured by Systems Innovation, Inc., Steam Hollow Road, R.D. 2 and Hallstead Plaza, Route 11, Hallstead, Pennsylvania.

   (1)  Required equipment. The following equipment or an equivalent substitute is required for calibration:

     (i)   Test Cable—Model TK 100-01. Manufactured by Systems Innovation, Inc. (See Appendix A, Figure 4 for an equivalent substitute.)

     (ii)   Power supply. Voltage output—11 to 14 volts dc .5 amp. (See Appendix A, Figure 5).

     (iii)   Frequency counter. Data Precision—Model 5740 or equivalent. (See Appendix A, Figure 5).

     (iv)   Digital delay generator. Range: .1 ms to 800 ms; Accuracy: + or - .1%. (See Appendix A, Figure 6).

   (2)  Calibration procedure. The calibration procedure shall be as follows:

     (i)   Interconnect the ESP Model TK 100 as shown in Appendix A, Figure 6.

     (ii)   Place the ESP switches in the following positions:

       (A)   Power On/Off—Off.

       (B)   Auto/Lock—Auto.

       (C)   Start On 1/On 2—On 1.

       (D)   Alert On/Off—Either one.

       (E)   Set Alert—on or off.

     (iii)   Set the digital delay generator to the nearest .1 ms—.0001 sec.—using the following formula:

 Delay time = 4091 divided by SPEED mph

       Example: To check the ESP Accuracy at 50 mph, set the delay time as follows:

 Delay Time = 4091 divided by 50 mph = 81.8 ms.

     (iv)   Place the ESP Power On/Off Switch to On. The ESP shall read 000.

     (v)   Check that the frequency counter reads 1,000,000 Hz + or -999 Hz.

     (vi)   Press the start switch and observe that the ESP reads within 1 mph of the desired test speed.

     (vii)   Check and record the ESP readings from 20 mph to 140 mph in 5 mph increments.

     (viii)   If the ESP failed one or more of the previous tests, it shall be returned to the manufacturer or to a service center approved by the manufacturer and tagged with the date tested, the serial number of the device and a notation of the general condition of the device.

     (ix)   If the ESP has passed the previous tests, calibration is complete.

 (c)  Speed Chek—(identified on the housing as Speed Chek model one, mfd. for: The Union Agency, Unionville, Pennsylvania 19375). Manufactured by Sterner Lighting Systems, Inc., 351 Lewis Avenue, Winsted, Minnesota.

   (1)  Required equipment. (See Appendix A, Figure 7 for test setup). The following equipment is required for calibration:

     (i)   Two pulse generators.

     (ii)   Dual channel oscilloscope or counter with interval capability.

     (iii)   Power supply—+5 volts.

   (2)  Calibration procedure. (See Appendix A, Figure 8 for interconnection diagram). Calibration procedures shall be as follows:

     (i)   A single circuit to generate the signals that simulate the front wheels of a vehicle crossing the tape switches shall be set up as set forth in Appendix A, Figures 5 and 6.

     (ii)   The elapsed time between the pulse on Outputs 1 and 2 is measured by the Speed Chek using the formula V x T = 3408. 3408 is the proper constant for a 5-foot roadswitch spacing.

     V = velocity in mph

     T = time in ms

     (iii)   The elapsed interval time is computed by the equation:

Web Only Graphic

     and compared to the readout on the Speed Chek.

     (iv)   If the readout on the face of the Speed Chek differs by more than 1 mph, the unit shall be returned to the manufacturer or to a service center approved by the manufacturer and tagged with the date tested, the serial number of the device and a notation of the general condition of the device.

 (d)  Model TK 100, Excessive Speed Preventor (ESP) Infrared—Manufactured by Systems Innovation, Inc., Steam Hollow Road, R. D. 2 and Hallstead Plaza, Route 11, Hallstead, Pennsylvania.

   (1)  Required equipment. The following equipment or an equivalent substitute is required for calibration:

     (i)   Test Cable—Model TK 100-01. Manufactured by Systems Innovation, Inc. (See Appendix A, Figure 4 for an equivalent substitute.)

     (ii)   Power supply. Voltage output—11 to 14 volts dc .5 amp.

     (iii)   Frequency counter. Data Precision—Model 5740 or equivalent.

     (iv)   Digital delay generator. Range: .1 ms to 800 ms; Accuracy: + or - .1%.

   (2)  Calibration procedure. The calibration procedure shall be as follows:

     (i)   Interconnect the ESP Model TK 100 Infrared as shown in Appendix A, Figure 6.

     (ii)   Place the ESP Infrared switches in the following positions:

       (A)   Power On/Off—Off.

       (B)   Auto/Lock—Auto.

       (C)   Start On 1/On 2—On 1.

       (D)   Alert On/Off—either one.

       (E)   Set Alert—on or off.

     (iii)   Set the digital delay generator to the nearest .1 ms—.0001 sec.—using the following formula:

 Delay time = 2045 divided by speed mph

 Example: To check the ESP Infrared Accuracy at 50 mph, set the delay time as follows:

 Delay time = 2045 divided by 50 mph = 40.9 ms.

     (iv)   Place the ESP Infrared Power On/Off Switch to On. The ESP Infrared shall read 000.

     (v)   Check that the frequency counter reads 1,000,000 Hz + or -999 Hz.

     (vi)   Press the start switch and observe that the ESP Infrared reads within 1 mph of the desired test speed.

     (vii)   Check and record the ESP Infrared readings from 20 mph to 140 mph in 5 mph increments.

     (viii)   Connect infrared sensors and turn on test panel switch.

     (ix)   Turn rotary switch to the 20 mph position and press test switch. Reading shall be within the tolerances of + 0 to -2 mph. Repeat for test speeds of 60 mph and 140 mph.

     (x)   If the ESP Infrared failed one or more of the previous tests, it shall be returned to the manufacturer or to a service center approved by the manufacturer and tagged with the date tested, the serial number of the device and a notation of the general condition of the device.

     (xi)   If the ESP Infrared has passed the previous tests, calibration is complete.

 (e)  Enradd, Model EJU-91—Manufactured by Y.I.S., Incorporated, 1049 Hartley Street, P. O. Box 3044, York, Pennsylvania 17404.

   (1)  Required equipment. The following equipment or an equivalent substitute is required for calibration:

     (i)   Two pulse generators.

     (ii)   Dual channel oscilloscope or frequency counter with interval capability.

     (iii)   Power supply, + 5 volts.

   (2)  Calibration procedure. (See Appendix A, Figure 8 for interconnection diagram). Calibration procedures shall be as follows:

     (i)   A single circuit to generate the signals that simulate the front wheels of a vehicle crossing the road sensor shall be set up as set forth in Appendix A, Figure 7.

     (ii)   The elapsed time between the pulse on Outputs 1 and 2 is measured by the Enradd using the formula V x T = 3408. 3408 is the proper constant for a 5-foot timing strip spacing.

 V = velocity in mph

 T = time in ms

     (iii)   The elapsed interval time is computed by:

       (A)   Calculation of the equation:

 V = 3408

 T

       (B)   Comparison of the readout on the oscilloscope/frequency counter to the readout on the Enradd.

     (iv)   If the readout on the face of the Enradd differs by more than +/- 1 mph, the unit shall be returned to the manufacturer or to a service center approved by the manufacturer and tagged with the date tested, the serial number of the device and a notation of the general condition of the device.

 (f)  Equipment, timing strips and other devices. The calibration and testing of equipment, timing strips and other devices actually used with a particular electronic device—nonradar—shall be performed under specifications provided by the manufacturer of an approved device using equipment specified by the manufacturer or equivalent substitute equipment.

Source

   The provisions of this §  105.56 adopted December 15, 1978, effective December 16, 1978, 8 Pa.B. 3578; amended May 18, 1979, effective May 19, 1979, 9 Pa.B. 1632; amended November 2, 1979, effective November 3, 1979, 9 Pa.B. 3652; amended December 28, 1979, effective December 29, 1979, 9 Pa.B. 4273; republished and compiled August 1, 1980, effective August 2, 1980, 10 Pa.B. 3199; amended August 22, 1986, effective August 23, 1986, 16 Pa.B. 3119; readopted December 4, 1987, effective December 5, 1987, 17 Pa.B. 5051; amended February 17, 1989, effective February 18, 1989, 19 Pa.B. 640; amended August 9, 1996, effective August 10, 1996, 26 Pa.B. 3833. Immediately preceding text appears at serial pages (218221) to (218224).

Cross References

   This section cited in 67 Pa. Code §  105a.4 (relating to system and speed timing device use and certification).



No part of the information on this site may be reproduced for profit or sold for profit.


This material has been drawn directly from the official Pennsylvania Code full text database. Due to the limitations of HTML or differences in display capabilities of different browsers, this version may differ slightly from the official printed version.