Pennsylvania Code & Bulletin
COMMONWEALTH OF PENNSYLVANIA

• No statutes or acts will be found at this website.

The Pennsylvania Code website reflects the Pennsylvania Code changes effective through 53 Pa.B. 8238 (December 30, 2023).

25 Pa. Code § 228.71. Timer.

§ 228.71. Timer.

 (a)  The control panel shall have a timer that is graduated in minutes and fractions of minutes or seconds. The timer shall have a preset time selector and an elapsed time indicator.

 (b)  The timer shall be cumulative and activated only during irradiation and shall retain its reading after irradiation is interrupted or terminated.

 (c)  The timer shall terminate irradiation when a preselected time has elapsed if the dose monitoring systems fail to do so.

Source

   The provisions of this §  228.71 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.



No part of the information on this site may be reproduced for profit or sold for profit.


This material has been drawn directly from the official Pennsylvania Code full text database. Due to the limitations of HTML or differences in display capabilities of different browsers, this version may differ slightly from the official printed version.