Pennsylvania Code & Bulletin
COMMONWEALTH OF PENNSYLVANIA

• No statutes or acts will be found at this website.

The Pennsylvania Code website reflects the Pennsylvania Code changes effective through 54 Pa.B. 488 (January 27, 2024).

25 Pa. Code § 240.606. QA requirements for laboratories.

§ 240.606. QA requirements for laboratories.

 (a)  CRMs for laboratories.

   (1)  Calibration. Each Department-listed CRM must have a current calibration. To have a current calibration, the CRM shall be calibrated in a Department-approved calibration facility within 1 year from the date of the previous calibration and when alterations or repairs are made to the CRM. A current calibration certificate shall be retained for each monitor. Analysis may not be performed on a monitor that was not calibrated during any portion of the testing period.

   (2)  Data collection log. CRM data shall be tracked on a form that contains all of the following:

     (i)   The CRM serial number.

     (ii)   The exposure dates and times.

     (iii)   The test result.

     (iv)   The address of the building tested.

     (v)   The test location in the building.

     (vi)   The name of the tester who placed the CRM.

     (vii)   The name of the tester who retrieved the CRM.

     (viii)   The calibration, repair and Department listing dates.

 (b)  CWLM for laboratories.

   (1)  Calibration. Each Department-listed CWLM must have a current calibration. To have a current calibration, the CWLM shall be calibrated in a Department-approved calibration facility within 1 year from the date of the previous calibration and when alterations or repairs are made to the CWLM. A current calibration certificate shall be retained for each monitor. Analysis may not be performed on a monitor that was not calibrated during any portion of the testing period.

   (2)  Data collection log. CWLM data shall be tracked on a form that contains all of the following:

     (i)   The CWLM serial number.

     (ii)   The exposure dates and times.

     (iii)   The test result.

     (iv)   The address of the building tested.

     (v)   The test location in the building.

     (vi)   The name of the tester who placed the CWLM.

     (vii)   The name of the tester who retrieved the CWLM.

     (viii)   The calibration, repair and Department listing dates.

 (c)  Electret ion chamber for laboratory analysis.

   (1)  Calibration. Each Department-listed electret reader shall have a current calibration. To have a current calibration, the electret reader shall be calibrated in a Department-approved calibration facility within 1 year from the date of the previous calibration and when alterations or repairs are made to the electret reader. Each electret reader shall be calibrated simultaneously with its corresponding reference electret’s recertification.

   (2)  Voltmeter routine instrument checks.

     (i)   Proper operation of the surface voltmeter shall be monitored following the manufacturer’s procedures for zeroing the voltmeter and analyzing the reference electrets.

     (ii)   A voltage reading of a reference electret difference of more than 2 volts from its specified value shall be considered a wrong reading and corrective action shall be taken.

     (iii)   If the voltmeter displays more than (±) 3 volts, corrective action shall be taken.

     (iv)   Voltmeter checks shall be conducted at least once each week while the voltmeter is in use and shall be documented. Documentation of routine instrument checks must include all of the following:

       (A)   The reader serial number.

       (B)   The date of analysis.

       (C)   Zero value.

       (D)   The reference electret values.

       (E)   Corrective actions performed.

   (3)  Known exposure measurements (spikes).

     (i)   Spikes shall be conducted at a rate of 3 for each 100 test devices deployed, with a minimum of 3 spikes for each certification year when tests were conducted in the certification year, and with a maximum of 6 spikes each month.

     (ii)   Spikes shall be analyzed in the same manner as all other testing.

     (iii)   Spikes shall be monitored using a means control chart. The means control chart must be established as follows:

       (A)   Using an RPE value of plus and minus 10%, which corresponds to the 1 sigma level.

       (B)   A warning level of the RPE of plus and minus 20%, which corresponds to the 2 sigma warning level.

       (C)   Control limits of the RPE of plus and minus 30%, which correspond to the 3 sigma control level.

     (iv)   Each RPE value shall be plotted on the means control chart within 1 week of return of the device from the radon chamber. If the RPE value is outside the 3 sigma control level, all measurements shall cease until the problem is evaluated and corrected. All evaluations shall be documented.

     (v)   In addition to the means control chart, all spikes shall be documented on a form that contains all of the following:

       (A)   The radon chamber name.

       (B)   The electret serial numbers.

       (C)   The RV from the radon chamber.

       (D)   The measured spike value or values.

       (E)   The individual RPE results.

       (F)   The certification year beginning date and end date.

       (G)   The exposure dates.

   (4)  Duplicate measurements.

     (i)   Duplicates shall be made in at least 10% of the total number of test devices deployed each month, or 50 each month, whichever is smaller.

     (ii)   The RPD shall be calculated for all duplicate results with an average of greater than or equal to 2.0 pCi/L. Two control charts shall be constructed to monitor duplicate precision:

       (A)   One for duplicates when the average is greater than or equal to 4.0 pCi/L.

       (B)   One for duplicates when the average is greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L.

     (iii)   Each RPD value shall be plotted on the control chart within 1 week of performing the duplicate measurement.

     (iv)   The RPD shall be tracked using control charts from ‘‘Protocols for Radon and Radon Decay Product Measurements in Homes,’’ EPA 402-R-92-003, May 1993, Appendix B, Exhibits B-2 and B-3.

     (v)   For duplicates when the average is greater than or equal to 4.0 pCi/L, all of the following apply:

       (A)   The control level shall be set at an RPD of 14%.

       (B)   The warning level shall be set at an RPD of 28%.

       (C)   The control limit shall be set at an RPD of 36%.

     (vi)   For duplicates when the average is greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L, all of the following apply:

       (A)   The control level shall be set at an RPD of 25%.

       (B)   The warning level shall be set at an RPD of 50%.

       (C)   The control limit shall be set at an RPD of 67%.

     (vii)   If the plotted RPD result falls outside of the control limit, the measurements shall cease until the problem is identified and corrected.

     (viii)   If the plotted RPD result falls outside of the warning level, ‘‘Protocols for Radon and Radon Decay Product Measurements in Homes,’’ EPA 402-R-92-003, May 1993, Appendix B, Exhibit B-5, shall be used to determine the action to be taken.

     (ix)   Documentation of duplicates must include all of the following:

       (A)   The device serial numbers.

       (B)   The exposure dates.

       (C)   Each duplicate measurement result.

       (D)   The RPD results.

 (d)  AC and LS.

   (1)  Calibration. All AC or LS laboratory systems shall be calibrated at least once every 12 months, when alterations or repairs are made to the system, or when a new batch of charcoal is received. This requires a determination of calibration factors for AC and LS devices by the exposure of these devices to a known concentration of radon in a Department-approved radon chamber. Calibration factors shall be determined for a range of exposure times and humidity levels.

   (2)  Laboratory control devices. The laboratory background level for each batch of AC and LS devices shall be established by each laboratory. Laboratories shall measure the background of at least 5% of unexposed AC and LS devices that have been processed according to their standard operating procedures (laboratory blanks).

   (3)  Routine counting system checks. Daily counting of a reference source shall be performed and documented. The characteristics of the check source (geometry, type of radiation emitted, and the like) must be similar to the samples to be analyzed. The count rate of the check sources must be high enough to yield reliable counting statistics in a short period of time, such as 1,000 to 10,000 counts per minute, to provide a maximum random uncertainty of 5%.

   (4)  Known exposure measurements (spikes).

     (i)   Spikes shall be conducted at a rate of 3 for each 100 test devices deployed, with a minimum of 3 spikes for each certification year when tests were conducted in the certification year, and with a maximum of 6 spikes each month.

     (ii)   Spikes shall be analyzed in the same manner as all other testing.

     (iii)   Spikes shall be monitored using a means control chart. The means control chart must be established as follows:

       (A)   Using an RPE value of plus and minus 10%, which corresponds to the 1 sigma level.

       (B)   A warning level of the RPE of plus and minus 20%, which corresponds to the 2 sigma warning level.

       (C)   Control limits of the RPE of plus and minus 30%, which correspond to the 3 sigma control level.

     (iv)   Each RPE value shall be plotted on the means control chart within 1 week of receiving the result from the laboratory. If the RPE value is outside the 3 sigma control level, all measurements shall cease until the problem is evaluated and corrected. All evaluations shall be documented.

     (v)   In addition to the means control chart, all spikes shall be documented on a form that contains all of the following:

       (A)   The radon chamber name.

       (B)   The device serial numbers.

       (C)   The RV from the radon chamber.

       (D)   The measured spike value or values.

       (E)   The individual RPE results.

       (F)   The certification year beginning date and end date.

       (G)   The exposure dates.

   (5)  Duplicate measurements.

     (i)   Duplicates shall be made in at least 10% of the total number of test devices deployed each month, or 50 each month, whichever is smaller.

     (ii)   The RPD shall be calculated for all duplicate results with an average of greater than or equal to 2.0 pCi/L. Two control charts shall be constructed to monitor duplicate precision:

       (A)   One for duplicates when the average is greater than or equal to 4.0 pCi/L.

       (B)   One for duplicates when the average is greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L.

     (iii)   Each RPD value shall be plotted on the control chart within 1 week of performing the duplicate measurement.

     (iv)   The RPD shall be tracked using control charts from ‘‘Protocols for Radon and Radon Decay Product Measurements in Homes,’’ EPA 402-R-92-003, May 1993, Appendix B, Exhibits B-2 and B-3.

     (v)   For duplicates when the average is greater than or equal to 4.0 pCi/L, all of the following apply:

       (A)   The control level shall be set at an RPD of 14%.

       (B)   The warning level shall be set at an RPD of 28%.

       (C)   The control limit shall be set at an RPD of 36%.

     (vi)   For duplicates when the average is greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L, all of the following apply:

       (A)   The control level shall be set at an RPD of 25%.

       (B)   The warning level shall be set at an RPD of 50%.

       (C)   The control limit shall be set at an RPD of 67%.

     (vii)   If the plotted RPD result falls outside of the control limit, the measurements shall cease until the problem is identified and corrected.

     (viii)   If the plotted RPD result falls outside of the warning level, ‘‘Protocols for Radon and Radon Decay Product Measurements in Homes,’’ EPA 402-R-92-003, May 1993, Appendix B, shall be used to determine the action to be taken.

     (ix)   Documentation of duplicates must include all of the following:

       (A)   The device serial numbers.

       (B)   The exposure dates.

       (C)   Each duplicate measurement result.

       (D)   The RPD results.

 (e)  ATs.

   (1)  Calibration. All AT laboratory systems shall be calibrated at least once every 12 months, when alterations or repairs are made to the system, or when each new batch or sheet of detector material is received. This requires a determination of calibration factors for AT devices by the exposure of these devices to different concentrations of radon in a Department-approved radon chamber.

   (2)  Laboratory control detectors. Laboratory control detectors for each batch of ATs shall be established and documented. Each laboratory shall measure the background of a statistically significant number of unexposed ATs. The laboratory control background value shall be subtracted from the field readings to produce a final result.

   (3)  Known exposure measurements (spikes).

     (i)   Spikes shall be conducted at a rate of 3 for each 100 test devices deployed, with a minimum of 3 spikes for each certification year when tests were conducted in the certification year, and with a maximum of 6 spikes each month.

     (ii)   Spikes shall be analyzed in the same manner as all other testing. The RV of a spike may not be revealed to the laboratory prior to analysis.

     (iii)   Spikes shall be monitored using a means control chart. The means control chart must be established as follows:

       (A)   Using an RPE value of plus and minus 10%, which corresponds to the 1 sigma level.

       (B)   A warning level of the RPE of plus and minus 20%, which corresponds to the 2 sigma warning level.

       (C)   Control limits of the RPE of plus and minus 30%, which correspond to the 3 sigma control level.

     (iv)   Each RPE value shall be plotted on the means control chart within 1 week of receiving the result from the laboratory. If the RPE value is outside the 3 sigma control level, all measurements shall cease until the problem is evaluated and corrected. All evaluations shall be documented.

     (v)   In addition to the means control chart, all spikes shall be documented on a form that contains all of the following:

       (A)   The radon chamber name.

       (B)   The device serial numbers.

       (C)   The RV from radon chamber.

       (D)   The measured spike value or values.

       (E)   The individual RPE results.

       (F)   The certification year beginning date and end date.

       (G)   The exposure dates.

   (4)  Duplicate measurements.

     (i)   Duplicates shall be made in at least 10% of the total number of test devices deployed each month, or 50 each month, whichever is smaller.

     (ii)   The RPD shall be calculated for all duplicate results with an average of greater than or equal to 2.0 pCi/L. Two control charts shall be constructed to monitor duplicate precision:

       (A)   One for duplicates when the average is greater than or equal to 4.0 pCi/L.

       (B)   One for duplicates when the average is greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L.

     (iii)   Each RPD value shall be plotted on the control chart within 1 week of performing the duplicate measurement.

     (iv)   The RPD shall be tracked using control charts from ‘‘Protocols for Radon and Radon Decay Product Measurements in Homes,’’ EPA 402-R-92-003, May 1993, Appendix B, Exhibits B-2 and B-3.

     (v)   For duplicates when the average is greater than or equal to 4.0 pCi/L, all of the following apply:

       (A)   The control level shall be set at an RPD of 14%.

       (B)   The warning level shall be set at an RPD of 28%.

       (C)   The control limit shall be set at an RPD of 36%.

     (vi)   For duplicates when the average is greater than or equal to 2.0 pCi/L and less than 4.0 pCi/L, all of the following apply:

       (A)   The control level shall be set at an RPD of 25%.

       (B)   The warning level shall be set at an RPD of 50%.

       (C)   The control limit shall be set at an RPD of 67%.

     (vii)   If the plotted RPD result falls outside of the control limit, the measurements shall cease until the problem is identified and corrected.

     (viii)   If the plotted RPD result falls outside of the warning level, ‘‘Protocols for Radon and Radon Decay Product Measurements in Homes,’’ EPA 402-R-92-003, May 1993, Appendix B, Exhibit B-5, shall be used to determine the action to be taken.

     (ix)   Documentation of duplicates shall include all of the following:

       (A)   The device serial numbers.

       (B)   The exposure dates.

       (C)   Each duplicate measurement result.

       (D)   The RPD results.

APPENDIX A


Radon Certification Fee Schedule



Testing Individual $525 every 2 years
Testing Employee $150 every 2 years
Testing Firm $1,050 every 2 years
Mitigation Individual $450 every 2 years
Mitigation Firm $1,050 every 2 years
Laboratory Individual $600 every 2 years
Laboratory Firm $1,125 every 2 years
Primary Testing Device Listing $150 every 2 years (1)
Course Provider $565 every 2 years (2)
Late Application Renewal $150
Late 45-Day Reporting $150 (3)
Radon Mitigation System Fee $50 for each radon mitigation system  installed or activated (4)

 The Department will review the adequacy of the fees established in this schedule at least once every 3 years and provide a written report to the EQB. The report must identify any disparity between the amount of program income generated by the fees and the costs to administer these programs, and must contain recommendations to increase fees to eliminate the disparity, including recommendations for regulatory amendments to increase program fees.

   (1)  Primary radon testers shall submit the Primary Testing Device Fee as specified in the Radon Certification Fee Schedule for each device they read or analyze, or both.

   (2)  A person approved by the Department to provide initial or continuing, or both, education courses shall submit the Course Provider Fee as specified in this appendix.

   (3)  Anyone not submitting the required 45-day testing or mitigation, or both, reporting within 90 days of the completion of the testing or mitigation, or both, activity (or if no activities have been performed during this period of informing the Department of same in writing) will be subject to the Late 45-Day Reporting Fee as specified in this appendix.

   (4)  The Department will waive the radon mitigation system fee for a local government employee or school employee who installs an active radon mitigation system in a school or local government building or activates a passive radon mitigation system in a school or local government building if the employee installs or activates the system pursuant to the employee’s official duties and the employee is not compensated for this service except through the employee’s salary.

Authority

   The provisions of this Appendix A amended under sections 301, 302 and 401 of the Radiation Protection Act (35 P.S. § §  7110.301, 7110.302 and 7110.401); section 1920-A of The Administrative Code of 1929 (71 P.S. §  510-20); and sections 8, 12 and 13 of the Radon Certification Act (63 P.S. § §  2008, 2012 and 2013).

Source

   The provisions of this Appendix A amended October 20, 2017, effective October 21, 2017, 47 Pa.B. 6482. Immediately preceding text appears at serial page (344628).

Cross References

   This appendix cited in 25 Pa. Code §  240.102 (relating to prerequisites for radon testing certification); 25 Pa. Code §  240.104 (relating to application filing deadline); 25 Pa. Code §  240.113 (relating to radon mitigation application contents); 25 Pa. Code §  240.114 (relating to application filing deadline); 25 Pa. Code §  240.123 (relating to radon laboratory application contents); 25 Pa. Code §  240.124 (relating to application filing deadline); 25 Pa. Code §  240.133 (relating to certification application contents); 25 Pa. Code §  240.204 (relating to certification renewal); and 25 Pa. Code §  240.309 (relating to radon mitigation system fee).

APPENDIX B


Non-interference Agreement for Real Estate Radon Testing

   Property name:
Property address:
Property city, state, zip:
Dates of test:

 I hereby agree to abide by the following conditions to ensure a valid radon test result:

 1) I will maintain closed-house conditions during the entire test period, and for 12 hours prior to any test of less than 96 hours, by doing the following:

 • Continuing normal operation of permanently installed HVAC systems.

 • Minimizing operation of dryers, range hoods, bathroom fans and other mechanical systems, understanding that drawing air out of the building may adversely affect the test results.

 • In buildings having permanently installed radon mitigation systems, keeping the mitigation system functioning during the testing interval.

 • Operating window air conditioning systems if set to recycle interior air.

 • Keeping all windows closed.

 • Keeping all external doors closed except for normal entry and exit.

 • Not operating whole-house fans. Removing portable window fans from the window or covering and sealing the window fan.

 • Not operating fireplaces, wood/coal stoves or combustion appliances, except water heaters and cooking appliances, unless they are the primary sources of heat for the building.

 • Not operating ceiling fans, portable dehumidifiers, portable humidifiers, portable air filters and window air conditioners within 20 feet of the detector.

 2) I will not interfere with or move the radon test device.

 If the certified tester determines that these conditions were not maintained, this test will be deemed invalid.

   


 
 

 Signature of Person     Printed Name of Person    Date
 in Control of Property    in Control of Property

Authority

   The provisions of this Appendix B issued under sections 12 and 13 of the Radon Certification Act (63 P.S. § §  2012 and 2013); section 302 of the Radiation Protection Act (35 P.S. §  7110.302); and section 1920-A of the Administrative Code (71 P.S. §  510.20).

Source

   The provisions of this Appendix B adopted October 26, 2018, effective January 24, 2019, 48 Pa.B. 6791.

APPENDIX C


Radon Exposure Tracking Record

 Name


  Month(s)

 Company Name


 Employee ID Number


  Year



Date Job Site Radon Level (pCi/L) Working
Level (WL)
Hrs. of
Exposure
Working
Level
Month
(WLM)
Cumulative Exposure(1) (WLM) Method used
to assess Exposure(2)




÷200=
x
÷170=





÷200=
x
÷170=





÷200=
x
÷170=





÷200=
x
÷170=





÷200=
x
÷170=





÷200=
x
÷170=





÷200=
x
÷170=





÷200=
x
÷170=





÷200=
x
÷170=





÷200=
x
÷170=





÷200=
x
÷170=





÷200=
x
÷170=





÷200=
x
÷170=





÷200=
x
÷170=


 1. Based upon an annual recommended health and safety limit of 4 working level months (4 WLM)

 2. Highest Premitigation Level (a) or On-site Measurement (b)

 WL = (pCi/L)/200 (assuming 50% ER)

Authority

   The provisions of this Appendix C issued under sections 12 and 13 of the Radon Certification Act (63 P.S. § §  2012 and 2013); section 302 of the Radiation Protection Act (35 P.S. §  7110.302); and section 1920-A of the Administrative Code (71 P.S. §  510.20).

Source

   The provisions of this Appendix C adopted October 26, 2018, effective January 24, 2019, 48 Pa.B. 6791.

Cross References

   This appendix cited in 25 Pa. Code §  240.305 (relating to health and safety program).



No part of the information on this site may be reproduced for profit or sold for profit.


This material has been drawn directly from the official Pennsylvania Code full text database. Due to the limitations of HTML or differences in display capabilities of different browsers, this version may differ slightly from the official printed version.